Publication:
Full geant4 and FLUKA simulations of an e-LINAC for its use in particle detectors performance tests

Thumbnail Image

Organizational Units

Authors

Piliçer, Ercan

Authors

Alpat, Behçet
Servoli, Leonello
Buono, Eduardo
Di Capua, Francesco
Menichelli, Mauro
Tucceri, Paola
Italiani, Marco

Advisor

Language

Publisher:

IOP Publishing

Journal Title

Journal ISSN

Volume Title

Abstract

In this work we present the results of full Geant4 and FLUKA simulations and comparison with dosimetry data of an electron LINAC of St. Maria Hospital located in Terni, Italy. The facility is being used primarily for radiotherapy and the goal of the present study is the detailed investigation of electron beam parameters to evaluate the possibility to use the e-LINAC (during time slots when it is not used for radiotherapy) to test the performance of detector systems, in particular those designed to operate in space. The critical beam parameters are electron energy, profile and flux available at the surface of device to be tested. The present work aims to extract these parameters from dosimetry calibration data available at the e-LINAC. The electron energy ranges from 4 MeV to 20 MeV. The dose measurements have been performed by using an Advanced Markus Chamber which has a small sensitive volume.

Description

Source:

Keywords:

Keywords

Instruments & instrumentation, Dosimetry concepts and apparatus, Performance of high energy physics detectors, Simulation methods and programs, Particle tracking detectors (solid-state detectors)

Citation

Alpat, B. vd. (2012). "Full geant4 and FLUKA simulations of an e-LINAC for its use in particle detectors performance tests". Journal of Instrumentation, 7,1-9.