Publication: Determination of full energy peak efficiency of HpGe detector from 59.5 to 1332.5 keV
Date
Authors
Gültekin, Ali Haydar
Kaynak, Gökay
Gürler, Orhan
Authors
Advisor
Language
Type
Publisher:
Natl Inst Science Communication - Niscair
Journal Title
Journal ISSN
Volume Title
Abstract
To detect very low radio activities in the environmental samples, it is sometimes necessary to know the full energy peak efficiency of the detector. This work proposes a full energy peak efficiency simulation which represents adequately the high purity germanium coaxial detector responses in the energy range 59.5-1332.5 keV and at distances of 5, 10 and 25 cm.
Description
Source:
Keywords:
Keywords
Physics, Monte Carlo simulation, HpGe detector, Efficiency, Calibration, Samples, Simulation, Gamma-ray, Coincidence corrections, Monte-carlo method, Including self-absorption
Citation