Publication:
Investigation of the basic dimensional properties and biomechanical performance of artificial ACL grafts produced by 3-D braiding method

Thumbnail Image

Organizational Units

Authors

Turşucular, Ömer Fırat

Authors

Turşucular, Ömer Fırat

Advisor

Language

Publisher:

idd3

Journal Title

Journal ISSN

Volume Title

Abstract

In the current study, an investigation of basic dimensional properties and biomechanical performance of artificial ACL grafts produced by the 3-D braiding method was carried out. Artificial ACL grafts were produced using the 3-D braiding process. All 3-D braiding grafts were produced using various technical yarns for raw and bio-finishing grafts. Ethylene oxide (EtO) sterilization was applied to all grafts. The effects of yarn types, yarn counts, and 3-D braiding constructions on EtO sterilized raw and bio-finished artificial ACL grafts were examined. In conclusion; the tensile strength and thickness values of all 3-D braiding structured ACL grafts were suitable for artificial ligament grafts. Additionally, its grammage values were extremely light. The yarn type, yarn count, and 3-D braiding construction were the most effective process parameters on thickness, grammage, and tensile strength values. They can be improved by using x2-layered 3-D braiding structures, thicker yarn counts, and more than 1 core yarn in similar studies to be conducted in the future.

Description

Source:

Keywords:

Keywords

Dimensional properties, Biomechanical performance, Bio-finishing, ACL graft, 3-D braiding