Publication: Dosimetric control of dose distribution calculated in computerized treatment planning system
No Thumbnail Available
Date
2011-01-01
Authors
GÜRLER, ORHAN
Şahin, Sevim
Gürler, Orhan
Gözcü, Sema
Kurt, Meral
Şengül, Kansu
Altay, Ali
Kahraman Çetintaş, Sibel
Özkan, Lütfi
Authors
Journal Title
Journal ISSN
Volume Title
Publisher
Kare Publ
Abstract
OBJECTIVESThe purpose of this study is providing dosimetric control by comparing dose values calculated in computerized treatment planning system (CTPS) and measured by Linear 2D-Array and ion chambers of various volumes.METHODSFields measured by Farmer, Semiflex and PinPoint ion chambers were created in CTPS. Using Linear 2D-Array, measurements were done in open, isocentric, oblique and wedge filtered fields which created in CTPS. Dose values were compared.RESULTSPinPoint ion chamber measured top dose values whereas Farmer ion chamber measured low values. The most compatible results with CTPS were obtained by Semiflex ion chamber. However, in 2D-Array measurements, concordance was observed with CTPS in central axis, difference reached 20 % in build-up and penumbra regions.CONCLUSIONThe most compatible results with CTPS were obtained by Semiflex ion chamber. Reasons of discordance in 2D-Array measurements were found as; varible dose around build-up region, and not to obtain the exact superpose of profiles' penumbra region.
Description
Keywords
Computerized treatment planning system, Linear 2d-array, Quality assurence in radiotherapy, Science & technology, Life sciences & biomedicine, Oncology