Yayın: LC determination of atropine sulfate and scopolamine hydrobromide in pharmaceuticals
Tarih
Kurum Yazarları
Cevheroğlu, Şemsettin
Yazarlar
Ceyhan, Timurhan
Kartal, Murat
Altun, Mehmet Levent
Tülemiş, F
Danışman
Dil
Türü
Yayıncı:
Elsevier
Dergi Başlığı
Dergi ISSN
Cilt Başlığı
Özet
An accurate, simple, reproducible and sensitive method for the determination of atropine sulfate and scopolamine hydrobromide has been developed and validated. Atropine sulfate and scopolamine hydrobromide were separated using a mu Bondapack C-18 column by isocratic elution with flow rate 1.0 ml/min. The mobile phase composition was methanol, water, formic acid (165:35:1; v/v/v) and pH adjusted 8.3 with triethylamine. The samples were detected at 230 nm using photo-diode array detector. The linear range of detection for atropine sulfate (I) and scopolamine hydrobromide (II) were between 10.38 and 1038 mug/ml with a limit of quantification (LOQ) of 10.38, 10.00 and 1034 mug/ml with an LOQ of 10.00 mug/ml respectively. The linearity, range, peak purity, selectivity, system performance parameters, precision, accuracy, robustness and ruggedness for (I) and (II) were also shown acceptable values.
Açıklama
Kaynak:
Anahtar Kelimeler:
Konusu
Atropine sulfate, Hyoscyamine-atropine, Scopolamine hydrobromide, High-performance liquid chromatography, Validation, Liquid-chromatography
Alıntı
Ceyhan, T vd. (2001). "LC determination of atropine sulfate and scopolamine hydrobromide in pharmaceuticals". Journal of Pharmaceutical and Biomedical Analysis, 25(3-4),399-406.