The applicability of a DETATA sorbent to the preconcentration of lead and cadmium followed by the determination of these elements in urine using atomic absorption spectrometry with electrothermal atomization was demonstrated. After preconcentration by a factor of 10, the limits of detection were 0.01 and 0.2 mu g/L for cadmium and lead, respectively. The accuracy of the results was supported by the analysis of Seronorm Trace Elements Urine (Batch no. 101021).

Açıklama

Kaynak:

Anahtar Kelimeler:

Konusu

Absorption, Agricultural chemicals, Atomic absorption spectrometry, Atomic physics, Atomic spectroscopy, Atomization, Atoms, Body fluids, Cadmium, Chemical elements, Electric heating, Fluid mechanics, Hydraulics, Lead, Sorption, Spectrometers, Spectrometry, Spraying, Trace analysis, Trace elements, Atomic absorption (AA), Electrothermal atomization (ETA), Electrothermal atomization atomic absorption spectrometry (ETAAS), Limits of detection (LOD), Pre concentration, Absorption spectroscopy, Separation, Samples, Matrix