Publication:
Lead fractionation analysis in lipstick samples by inductively coupled plasma-mass spectrometry

No Thumbnail Available

Date

2015-01-01

Authors

Authors

Yılmazcan, Özlem
Erdemir, Ümran Seven
İzgi, Belgin
Özer, Elif Tümay
Güçer, Şeref

Journal Title

Journal ISSN

Volume Title

Publisher

Taylor & Francis

Research Projects

Organizational Units

Journal Issue

Abstract

A new analytical approach based on fractionation was introduced for lead in lipstick samples. Different separation techniques including n-hexane, glycerol extraction, and activated carbon adsorption were used to characterize the lipid fraction, polar and aromatic components of the samples. Additionally, artificial saliva and food stimulant extractions were used for the risk assessment studies. Trace metal levels in fractions were determined by inductively coupled plasma-mass spectrometry. Method validation parameters in the total element determinations were defined in terms of detection limits, accuracy, and precision. The limits of detection and quantification were 0.02 and 0.07mgkg(-1) for Pb; whereas the repeatability and reproducibility of the results based on percent relative standard deviation were 3.0% and 7.2% for lead, respectively.

Description

Keywords

Activated carbon, Adsorption, Elements, Cadmium, Copper, Milk, Fractionation, Icp-ms, Lead, Lipstick, Spectroscopy

Citation

Collections

1

Views

0

Downloads

Search on Google Scholar