Publication:
Dislodgement of a sirolimus-eluting stent in the circumflex artery and its successful deployment with a small-balloon technique

dc.contributor.buuauthorŞentürk, Tunay
dc.contributor.buuauthorYEŞİLBURSA, DİLEK
dc.contributor.buuauthorŞENTÜRK, TUNAY
dc.contributor.buuauthorÖZDEMİR, BÜLENT
dc.contributor.buuauthorÖzdemir, Bülent
dc.contributor.buuauthorSerdar, Osman Akın
dc.contributor.buuauthorSERDAR, OSMAN AKIN
dc.contributor.departmentTıp Fakültesi
dc.contributor.departmentKardiyoloji Ana Bilim Dalı
dc.contributor.orcid0000-0003-1744-8883
dc.contributor.researcheridAAF-5116-2019
dc.contributor.researcheridC-1517-2017
dc.date.accessioned2024-09-24T06:02:03Z
dc.date.available2024-09-24T06:02:03Z
dc.date.issued2011-07-01
dc.description.abstractCoronary stent dislodgement or embolization before deployment is a rare but serious complication in interventional cardiology. A 60-year-old male presented with unstable angina five years after coronary artery bypass surgery. There was a stenosis (70%) in the obtuse marginal branch of the circumflex artery. During percutaneous coronary intervention, a sirolimus-eluting stent was stripped from its balloon mainly because of significant proximal angulation and incarcerated within the proximal circumflex artery. A smaller balloon dilatation catheter was advanced and pushed through the inside of the slipped stent. Using this technique, the stent could be advanced into the lesion and was successfully deployed. To our knowledge, this is the first case report on sirolimus-eluting stent dislodgement.
dc.identifier.doi10.5543/tkda.2011.01248
dc.identifier.endpage421
dc.identifier.issn1016-5169
dc.identifier.issue5
dc.identifier.startpage418
dc.identifier.urihttps://doi.org/10.5543/tkda.2011.01248
dc.identifier.urihttps://hdl.handle.net/11452/45090
dc.identifier.volume39
dc.identifier.wos000421531500011
dc.indexed.wosWOS.ESCI
dc.language.isoen
dc.publisherTurkish Soc Cardiology
dc.relation.journalTurk Kardiyoloji Dernegi Arsivi-archives Of The Turkish Society Of Cardiology
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi
dc.rightsinfo:eu-repo/semantics/closedAccess
dc.subjectAngioplasty
dc.subjectBalloon
dc.subjectCoronary/instrumentation
dc.subjectEquipment failure
dc.subjectStents
dc.subjectScience & technology
dc.subjectLife sciences & biomedicine
dc.subjectCardiac & cardiovascular systems
dc.subjectCardiovascular system & cardiology
dc.titleDislodgement of a sirolimus-eluting stent in the circumflex artery and its successful deployment with a small-balloon technique
dc.typeArticle
dspace.entity.typePublication
local.contributor.departmentTıp Fakültesi/Kardiyoloji Ana Bilim Dalı
relation.isAuthorOfPublication4120305e-77c5-4175-a994-c743eddf6793
relation.isAuthorOfPublicationc653a7c1-4ead-46da-83e5-de7378a3726d
relation.isAuthorOfPublicationac4f107d-9807-4393-b0fb-db1d981f6490
relation.isAuthorOfPublication.latestForDiscovery4120305e-77c5-4175-a994-c743eddf6793

Files